Fluke 729 Pro Automatic Pressure Calibrator
Nøglefunktioner
- Automatic pressure generation and regulation to 1000psi
- Replaceable pressure control modules
- Built-in test templates, one-key calibration
- HART communication
- Pressure leakage testing
- Automatic fine-pressure adjustment, automatic pressure stabilization
- Powerful, replaceable Li-ion battery
- Quick pressure release valve with a muffler
- Extend measurement range with 750P Series pressure modules
- Measure temperature with 720RTD probe
- Bright dual-three-channel /color graphic display
- Weight <5kg
- Rugged and portable design with standard 3-year warranty
- Absolute and gage pressure modules up to 7MPa.
- Use the modules to change the Fluke 729Pro to a different output pressure range
Produktoversigt: Fluke 729 Pro Automatic Pressure Calibrator
Bærbar automatisk trykkalibrator forenkler trykkalibrering
Fluke 729Pro Automatisk Trykkalibrator er udviklet specielt med procesteknikere i tankerne for at forenkle trykkalibreringsprocessen og give hurtigere og mere nøjagtige testresultater. Teknikere ved, at kalibrering af tryk kan være en tidskrævende opgave, men 729Pro gør det nemmere end nogensinde før med en indbygget elektrisk pumpe, der giver automatisk trykgenerering og -regulering i en letanvendelig og robust bærbar pakke.
Automatisk trykgenerering og -kontrol
Med 729Pro kan du blot indtaste et tryk, hvorefter kalibratoren automatisk pumper til det ønskede indstillingspunkt. Med udskiftelige plug-in trykstyremoduler og trykgenerering og -regulering op til 1000 psi (70 bar, 7000 kPa). Derefter stabiliserer den interne finjusteringskontrol automatisk trykket til den ønskede værdi. Du kan også udfylde en testskabelon, hvorefter 729Pro automatisk pumpes til og dokumenterer en flerpunktskalibreringstest ved ét tryk på en knap.
Nem kalibreringsdokumentation
Fluke 729Pro kan automatisk teste flere trykprøvningspunkter og automatisk dokumentere resultaterne. 729Pro gør arbejdet for dig ved at dokumentere og markere resultater, der ligger uden for tolerance, i rødt på det grafiske display for:
- Anvendt tryk
- Målt mA, mV eller digital PV
- % fejl for hvert testpunkt
Upload og administrer dokumenterede kalibreringsresultater med DPCTrack2™ Calibration Management Software, hvilket gør det nemt at administrere din instrumentering, oprette planlagte tests og rapporter og administrere kalibreringsdata.
HART-kommunikation
Indbyggede HART-kommunikationsfunktioner muliggør HART-transmitter mA justeringer, HART-konfigurering og evnen til at justere til anvendte 0% og 100% værdier. Du kan også konfigurere forskellige opgaver såsom at ændre et transmitter tag, måleenheder og områdeskift. Andre understøttede HART-kommandoer inkluderer indstilling af faste mA udgange til fejlfinding, læsning af enhedskonfiguration og variabler og læsning af enhedsdiagnostik.
Robust, bærbart design
729Pro er testet og garanteret til at modstå en 1-meters faldtest, så den er klar til kalibreringsarbejde på feltinstrumentering. Med de to modeller – 4M op til 600 psi (40 bar, 4000 kPa) og 7M op til 1000 psi (70 bar, 7000 kPa) – at vælge imellem er Fluke 729Pro automatiske trykkalibratorer designet til at levere, præcis når du har brug for det. En standard middelfast bæretaske designet til feltarbejde med plads i tasken til opbevaring af testslanger, fittings, testledninger og små håndværktøjer betyder, at det er let at bære alt, hvad der kræves til felttrykkalibreringer.
For at opnå større nøjagtighed og opløsning ved kalibrering af lavt tryk giver Fluke 729Pro dig mulighed for at vælge og indsætte et af 6 Fluke trykmoduler for at udvide trykområdet fra 0 til 600 psi (40 bar, 4000 kPa).
Specifikationer: Fluke 729 Pro Automatic Pressure Calibrator
Key parameters | 729 Pro 7M | 729 Pro 4M |
Pressure control range | -13 to 1000 psi, -90 to 7000 kPa, -0.90 to 70 bar | -13 to 600 psi, -90 to 4000 kPa, -0.90 to 40 bar |
Pressure type | Gauge/absolute (Depends on the type of internal or external pressure module | |
Pressure accuracy and Control stability | 0.02% of FS, 0.005% of FS | |
Replaceable pressure measurement modules | Yes, 6 modules | Yes, 4 modules |
mA sourcing/measurement | Yes | |
Voltage measurement | Yes | |
Loop power | 24 V dc | |
Pressure switch testing | Yes | |
Pressure leakage testing | Yes | |
HART Communication | Yes, basic commands supported | |
Autostep / auto ramp | Yes (with customizable speed, mode and style) | |
Battery | Rechargeable lithium battery with charge indication (charged in the instrument or separately) | |
Language | English |
Electrical specification | |
All specifications are valid to 110% of range, except 24 mA source and simulate which are valid to 100% of range. | |
Ranges | |
mA measure, source and simulate | 0 mA to 24 mA |
Volts dc measurement | 0 V dc to 30 V dc |
Resolution | |
mA dc source, simulate and measure | 1 μA |
Voltage dc measurement | 1 mV |
Accuracy | 0.01% FS ±2 LSD all ranges (at 23 °C ±5 °C) |
Stability | 20 ppm FS /°C (-10 °C to +18 °C and 28 °C to 50 °C) |
mA simulation working voltage | 12 V dc to 30 V dc |
Loop compliance voltage | 24 V dc @ 20 mA |
Temperature measurement , 100 Ω Pt(385) RTD | -50 °C to 150 °C |
Temperature resolution | 0.01 °C |
Temperature accuracy | ±0.1 °C ±0.25 °C combined uncertainty when using 720 RTD probe (optional accessory) |
Drive capability (mA source) | 1200 Ω (without HART resistor), 950 Ω (with internal HART resistor) |
Mechanical specification | |
Dimension (HxWxL) | 290 x 215 x100 mm, (4.3 x 4.4 x 9.4 in) |
Weight | 4.9 kg, (10.8 lbs) |
Environmental specification | |
Operating temperature | -10 °C to 50 °C |
Pressure control | 0 °C to 50 °C |
Battery charging | 0 °C to 40 °C |
Storage temperature | -20 °C to 60 °C |
Operating altitude | <3000 m |
Storage altitude | <13000 m |
Operating humidity | Non-condensing (<10°C) ≤90 % RH (10 °C to 30 °C) ≤75 % RH (30 °C to 40 °C) ≤45 % RH (40 °C to 50 °C) |
Pressure Module Selection Table | ||||||
Pressure module | Pressure type | Range | Resolution | Overall uncertainty | 729Pro 4M | 729Pro 7M |
FLK-PMM-200K | Gauge | (-15 to 30) psi | 0.0001 psi | 0.02 % FS / year, maximum | • | • |
(-100 to 200) kPa | 0.001 kPa | |||||
(-1 to 2) bar | 0.00001 bar | |||||
FLK-PMM-200KA | Absolute | (0 to 30) psi | 0.0001 psi | 0.05 % FS / year, maximum | • | • |
(0 to 200) kPa | 0.001 kPa | |||||
(0 to 2) bar | 0.00001 bar | |||||
FLK-PMM-2000K | Gauge | (-13 to 300) psi | 0.001 psi | 0.02 % FS / year, maximum | • | • |
(-90 to 2000) kPa | 0.01 kPa | |||||
(-0.90 to 20) bar | 0.0001 bar | |||||
FLK-PMM-4000K | Gauge | (-13 to 600) psi | 0.01 psi | 0.02 % FS / year, maximum | • | • |
(-90 to 4000) kPa | 0.01 kPa | |||||
(-0.90 to 40) bar | 0.0001 ba | |||||
FLK-PMM-7000K | Gauge | (-13 to 1000) psi | 0.01 psi | 0.02 % FS / year, maximum | • | |
(-90 to 7000) kPa | 0.01 kPa | |||||
(-0.90 to 70) bar | 0.0001 ba | |||||
FLK-PMM-7000KA | Absolute | (0 to 1000) psi | 0.01 psi | 0.02 % FS / year, maximum | • | |
(0 to 7000) kPa | 0.01 kPa | |||||
(0 to 70) bar | 0.0001 ba | |||||
Dry air or non-corrosive gases only / 15 °C to 35 °C, all ranges |